B-24 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
etCO
2
Microstream Pod
Physical Attributes
Size (H x W x D):
Approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Weight:
Pod: 0.476 kg (1.05 lb)
Degree of protection
against ingress of water:
IPX1 (per IEC 60529)
Electrical Specifications
Power source:
Powered directly from monitor
Protection against
electric shock:
Pod electrically isolated by host device.
Degree of protection
against electric shock:
Type BF
Mode of operation:
Continuous
Defibrillation protection:
Defibrillation-Proof Applied Part per IEC 60601-1
Environmental requirements
Temperature range:
Operating: 0 to 50 °C (32 to 122 °F)
Storage: –20 to 70 °C (–4 to 158 °F)
Relative humidity:
Operating: 10 to 95%, non-condensing
Storage: 10 to 95% (with packaging)
Atmospheric pressure:
Operating: 430 to 795 mmHg (57 to 106 kPa)
Storage: 430 to 795 mmHg (57 to 106 kPa)
Regulatory Requirements
Standards: IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable sections
only, Oxygen analyzers for monitoring patient breathing mixtures);
EN ISO 21647:2009