MONITORING A CCESSORIES
VF8 DELTA/DELTA XL/KAPPA B-25
Hemodynamic Pods
User Interface
User Controls: Fixed keys (C.O. Start, IBP Zero, Wedge)
Displays: HemoMed:No LCDs
Hemo2: two 4-character LCDs.
Hemo4: four 4-character LCDs.
MPod – Quad Hemo: No LCDs
Connections: Hemo2: 2 invasive pressures, C.O., 2 temperatures, single
cable connecting pod to monitor
Hemo4, MPod – Quad Hemo, and HemoMed:
4 invasive pressures, C.O., 2 temperatures, single cable
connecting pod to monitor
Degree of protection against ingress
of water:
Hemo2, Hemo4, and HemoMed: IPX0 (per IEC 60529)
MPod – QuadHemo: IPX1 (protected against harmful
effects of dripping water) per IEC 60529
Physical Attributes
Size (H x W x D): Hemo2, Hemo4, and HemoMed:140 x 205 x 60 mm
(5.5 x 8.1 x 2.3 in)
MPod – QuadHemo: 205 x 110 x 80 mm (8.1 x 4.3 x 3.2 in)
Weight: Hemo2 and HemoMed:0.7 kg (1.6 lb.)
Hemo4: 0.9 kg (1.9 lb.)
MPod – Quad Hemo: 0.48 kg (1.1 lb)
NOTE: Hemo2 weight includes one transducer adapter
block. Hemo4 weight includes two transducer adapter
blocks. Both weights exclude mounting clamp.
For the MPod – Quad Hemo the weight includes four
transducer cables but excludes the mounting clamp and
rod.
Electrical Specifications
Power source: Powered directly from the monitor
Protection against electric shock: Type CF
Mode of operation: Continuous
Defibrillation protection: Per IEC 60601-2-34
Environmental Requirements
Temperature range: Operating:
Hemo2, Hemo4, HemoMed:10 to 40 °C (50 to 104 °F)
MPod – QuadHemo: 0°C to 45°C (32°F to 113°F)
Storage:
Hemo2, Hemo4, HemoMed: –20 to 50 °C (–4 to 122 °F)
MPod – QuadHemo: –40°C to +70°C (–40°F to +158°F)