2-6 DELTA/DELTA XL/KAPPA VF8
2 M
ONITOR SETUP
Max.
Channels
Determines the number of waveform
channels and parameters displayed.
Click on 4, 5, 6, 7, or 8
NOTE: The number of waveforms
depends on the software option you
have installed (see page 1-20 for a list
of options).
Bottom
Channel
Configures the bottom waveform channel
to display a waveform or 3 parameter
boxes.
Waveform, Parameters
ECG
Channels
Determines the number and format of
ECG waveforms displayed.
Click on the following settings:
ECG1 – Displays the primary ECG
waveform
ECG1 & 2 – Displays 2 waveforms
ECG1 & 2 & 3 – Displays 3
waveforms
Cascade – Cascades ECG1 data
into second channel
ARR
Monitoring
Selects the Arrhythmia mode
(For detailed information, see page 9-5.)
•Select OFF to disable arrhythmia
monitoring.
•Select FULL to enable full arrhythmia
monitoring.
•Select BASIC to enable basic
arrhythmia monitoring.
ST
Monitoring
Enables and disables ST monitoring (for
detailed information, refer to chapter 10).
•Select ON to enable ST monitoring.
•Select OFF to disable ST monitoring.
RESP
Monitoring
Enables and disables respiration
monitoring (for detailed information, refer
to chapter 12).
•Select ON to enable respiration
monitoring.
•Select OFF to disable respiration
monitoring.
Display
Mode
Reduces Main screen clutter by
displaying only parameters associated
with a connected device (see Parameter
Priority on page 2-5).
•Select Manual to display all
parameters and assign them a
priority on the Parameter Priority
screen.
•Select Automatic to display active
parameters only.
Agent
Display
Displays an anesthetic agent parameter
box.
•Select
ON to display agent.
•Select OFF to cancel display.
N2O Display Displays N
2
O/O
2
parameter box.
NOTE: This selection is ghosted unless
MultiGas or MultiGas+ module is
connected.
•Select ON to display N
2
O/O
2
.
•Select OFF to cancel display.