Dräger Oxylog - Demand Valve Readiness Check; Pre-Use Demand Valve Checks; Demand Valve Operation; Mask Fitting and Seal

Dräger Oxylog
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Check
the
demand
valve
for
operational
readiness
-
Prior to each use
Push a catheter adapter size 5 into the
patient port of the non-rebreathing
valve. Generate a suction with your
mouth:
0
Gas should begin to flow.
Stop sucking:
0
The gas flow is cut off.
Remove the catheter adapter.
Operation
Push the face mask onto the patient
port of the non-rebreathing valve and
make sure that the mask makes a
tight seal with the face.
Upon commencement of spontane-
ous breathing:
0
Set pneumatic main switch to 0.
Spontaneous breathing with posi-
tive airway pressure CPAP is not
possible!
Care
Stripping down
0
Detach ventilation hose from
Oxylog
*
0
Detach non-rebreathing valve from
hose.
0
Detach mask from non-rebreathing
valve.
0
Turn cover of non-rebreathing
valve anti-clockwise by 45and
remove.
0
Remove diaphragm from valve
housing.
Cleaning
Clean disassembled patient system
using water to which a detergent has
been added, and dry well.
Disinfection
The disinfected patient system can
be subjected to bath disinfection
e.g. using Tego 103 G (Messrs.
Th. Goldschmidt AG,
Essen)
and
observe the manufacturers instruc-
tions for use.
Sterilization
The valve housing of the
non-re-
breathing valve must not be sub-
jected to thermal disinfection.
The diaphragm, the cover and the
ventilation hose can be
autoclaved
at
134°C.
16