Page #16 background imageLoading...
Page #16 background image
Check
the
demand
valve
for
operational
readiness
-
Prior to each use
Push a catheter adapter size 5 into the
patient port of the non-rebreathing
valve. Generate a suction with your
mouth:
0
Gas should begin to flow.
Stop sucking:
0
The gas flow is cut off.
Remove the catheter adapter.
Operation
Push the face mask onto the patient
port of the non-rebreathing valve and
make sure that the mask makes a
tight seal with the face.
Upon commencement of spontane-
ous breathing:
0
Set pneumatic main switch to 0.
Spontaneous breathing with posi-
tive airway pressure CPAP is not
possible!
Care
Stripping down
0
Detach ventilation hose from
Oxylog
*
0
Detach non-rebreathing valve from
hose.
0
Detach mask from non-rebreathing
valve.
0
Turn cover of non-rebreathing
valve anti-clockwise by 45and
remove.
0
Remove diaphragm from valve
housing.
Cleaning
Clean disassembled patient system
using water to which a detergent has
been added, and dry well.
Disinfection
The disinfected patient system can
be subjected to bath disinfection
e.g. using Tego 103 G (Messrs.
Th. Goldschmidt AG,
Essen)
and
observe the manufacturers instruc-
tions for use.
Sterilization
The valve housing of the
non-re-
breathing valve must not be sub-
jected to thermal disinfection.
The diaphragm, the cover and the
ventilation hose can be
autoclaved
at
134°C.
16

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog and is the answer not in the manual?

Dräger Oxylog Specifications

General IconGeneral
Power SourceAC Power, Internal Battery
Peak FlowUp to 120 L/min
TypeVentilator
Operating ModesSIMV
Tidal Volume Range50 mL to 2000 mL
PEEP0 to 20 cm H2O
FiO221 to 100 %

Related product manuals