EasyManua.ls Logo

Dräger V500 - Key Concepts and Comparisons; Avea vs. Dräger Function Comparison

Dräger V500
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 | © 2015
Increase O
2
Function
Avea: FiO
2
increases by 10%
Increase FiO
2
Button
Dräger: FiO
2
increases by 1.3 times the set FiO
2
Example 1: Set FiO
2
is 25%. O
2
Suction is activated, FiO
2
will increase to 33%
Example 2: Set FiO
2
is 60%. O
2
Suction is activated, FiO
2
will increase to 78%.
O
2
Suction Button
o Push O
2
Suction on screen
o Confirm by pushing control knob
180 seconds of oxygen increase
Differences in Increase O
2
or
O
2
Suction