29 | © 2015
• Volume ventilation is a mode in which the ventilator will deliver a set volume
− Peak inspiratory pressures will vary depending on patient’s compliance and resistance
− Peak inspiratory pressures can be limited for safety
• Peak inspiratory pressure will increase:
− Decreased compliance
Surfactant deficiency
Inflammation/immaturity
− Increased resistance
Secretions
Kinked ET tube
Pneumothorax
• Peak inspiratory pressure will decrease:
− Increased compliance
Surfactant administration
− Decreased resistance
ET tube suctioned and secretions removed
Bronchodilator given