Instructions for use VarioLux 21
English
Recommended separation distances from wireless
communication devices
To ensure the full functional integrity of this device, there must be
a separation distance of at least 1.0 m (3.3 ft) between this device
and wireless communication devices.
Immunity against Test level and required electromagnetic environment
Electrostatic discharge (ESD) (IEC 61000-4-2) Contact discharge: ±8 kV
Air discharge: ±15 kV
Fast transient electrical disturbances (bursts) (IEC 61000-4-4) Power cable: ±2 kV
Longer signal input lines/output lines: ±1 kV
Impulse voltages (surges) (IEC 61000-4-5) Voltage, external conductor – external conductor: ±1 kV
Voltage, external conductor – protective ground conductor: ±2 kV
Magnetic fields at mains frequency (IEC 61000-4-8) 50 Hz: 30 A/m
Voltage dips and short interruptions in the supply voltage
(IEC 61000-4-11)
30 % to 100 %, 10 ms to 5 s, different phase angles
Radiated high-frequency disturbances (IEC 61000-4-3) 80 MHz to 2.7 GHz: 10 V/m
Conducted high-frequency disturbances (IEC 61000-4-6) 150 kHz to 80 MHz: 3 V, ISM bands and amateur radio bands: 6 V
Electromagnetic fields in the vicinity of wireless communication
devices
Various frequencies from 385 MHz to 5785 MHz: 9 V/m to 28 V/m