EasyManuals Logo

Dräger VarioLux User Manual

Dräger VarioLux
Go to English
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
English
20 Instructions for use VarioLux
After reprocessing
Preparation before next use of device
Refer to the "Assembly and preparation"chapter on page 15.
Checking the operational readiness
Prerequisites:
The device has been assembled and prepared so that it is
ready for operation.
Procedure:
1 Check the operational readiness, refer to the "Operation"
chapter on page 15.
Service
Definition of maintenance concepts
Inspection
Regular inspections must be carried out according to the following
specifications and at the specified intervals. Technical
documentation is available on request.
Carrying out the inspection
1 Check that the respective instructions for use are present.
2 Check the following functions in accordance with the
instructions for use:
Refer to the "Operation" chapter
3 Check that the product is in good condition:
All labels are complete and legible
There is no visible damage
4 Observe the instructions for use and check that all necessary
components and accessories are present.
5 Check the electrical safety in accordance with the IEC 62353
standard.
Repair
Dräger recommends that all repairs are performed
by DrägerService and that only authentic Dräger parts are used.
Disposal
Disposing of the device
The disposal of electrical and electronic devices is subject to
special guidelines.
This device must be disposed of in accordance with national
regulations. In countries of the European Union, Dräger will
organize the return of the device. Additional information is
available at www.draeger.com (search term: WEEE).
Technical data
EMC declaration
General information
This device was tested for electromagnetic compatibility using
accessories from the list of accessories. Other accessories may
only be used if they do not compromise the electromagnetic
compatibility. The use of non-compliant accessories may result in
increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
This device may be used in the direct vicinity of other devices only
if Dräger has approved this device arrangement. If no approval
has been given by Dräger, it must be ensured that this device
functions correctly in the desired arrangement before use. The
instructions for use for the other devices must be followed.
Electromagnetic environment
This device may only be used in environments specified in the
"Environments of use" section on page 14.
Concept Definition
Service All measures (inspection, maintenance, re-
pair) intended to maintain or restore the
functional integrity of a product
Inspection Measures intended to determine and as-
sess the current state of a product
Maintenance Regular specified measures intended to
maintain the functional integrity of a prod-
uct
Repair Measures intended to restore the function-
al condition of a product after a device mal-
function
Measure Interval User group
Inspection Every 2 years Specialized service
personnel
WARNING
This device may only be used adjacent to or stacked with
other devices if the configuration is approved by Dräger. If
adjacent or stacked use of non-approved configurations is
absolutely necessary, verify normal operation of the device
in this configuration before using the device.
In any case, strictly observe the instructions for use of the
other devices.
Emissions Compliance
Radiated emissions Class B, group 1
(30 MHz to 1 GHz)
Conducted emissions Class B, group 1
(150 kHz to 30 MHz)
Harmonic emissions Class A
Voltage fluctuation emissions
(flicker)
Requirement is met.