Procedural Silence indicator Indicates that the unit is in procedural silence. The procedural
silence interval is 5 min. During procedural silence, the low Baby
Temp alarms are blocked.
Silence/Reset key Has 3 functions: silences alarms, resets Manual Mode, and sets
procedural silence.
If an alarm sounds in Manual Mode, press to silence the High
Temp alarm for 2 min, reset the Chk Patient alarm, and silence
the Chk Patient alarm at any time after 10 min. If it is not
pressed, the heater turns off after 15 minutes; press to restore
heater power.
If an alarm sounds in Baby Mode, press to silence the High Temp
alarm for 2 min or the Baby Temp alarm for 10 min. During non-
alarm conditions, press to enter procedural silence.
Pre-Warm indicator Indicates that the controller is operating in Pre-Warm Mode.
Manual indicator Indicates that the controller is operating in Manual Mode.
Baby indicator Indicates that the controller is operating in Baby Mode.
Mode Select key Press to select Pre-Warm, Manual, or Baby Mode of operation.
NOTE:
Defaults to Pre-Warm upon power up.
Skin temperature probe connector Accepts the skin temperature probe for monitoring the infant’s
skin temperature. When connected, the Baby Temperature
display indicates the temperature sensed by the probe when in
Manual or Baby Modes. When in Pre-Warm Mode for software
version 1.08 or greater, the display indicates “PrE.” When the
probe is disconnected, the Baby Temperature display goes blank
in Manual or Baby Modes and indicates "PrE" in Pre-Warm
Mode. When disconnected in Baby Mode, the probe alarm also
occurs.
Keypad Lock key Disables keypad to avoid accidental changes to settings.
Press to disable the >37°C, Up arrow, Down arrow, and Mode
Select keys. The Keypad Lock key illuminates to indicate the
keypad is locked. Press again to enable the keys.
Examination Light key Press to turn the examination light located in the warmer module
on or off.