Technical Manual | X-act
®
7000 57
Troubleshooting | en
Remote pump - battery
low
The charge status of the
rechargeable battery for
the X-am pump is low
Promptly recharge the X-
am pump
Languages - not found The language table could
not be found in the mem-
ory
The X-act 7000 can still be
used, but only in English;
install the current lan-
guage table on the X-act
7000 with CC-Vision
Languages - unknown The version of the lan-
guage table is incompati-
ble with the X-act 7000
firmware
The X-act 7000 can still be
used, but only in English;
install the current lan-
guage table on the X-act
7000 with CC-Vision
Languages - corrupted The language table is
faulty
The X-act 7000 can still be
used, but only in English
Substances - not found The substance table could
not be found in the mem-
ory
The X-act 7000 can still be
used, but “Unknown gas”
is always displayed as the
gas name; install the cur-
rent substance table on
the X-act 7000 with CC-
Vision
Substances - corrupted Substance table is faulty The X-act 7000 can still be
used, but “Unknown gas”
is always displayed as the
gas name; install the cur-
rent substance table on
the device with CC-Vision
Databases - incompatible The version of the sub-
stance table is incompati-
ble with the X-act 7000
firmware
The X-act 7000 can still be
used, but “Unknown gas”
is always displayed as the
gas name; install the cur-
rent substance table on
the X-act 7000 with CC-
Vision
Locations - corrupted The measurement location
table is faulty
Reconfigure the measure-
ment locations with CC-
Vision
Memory - exceeded The memory of the X-act
7000 is almost exhausted
If possible, remove unnec-
essary measurement loca-
tions from the
measurement location
table with CC-Vision
Warnings Cause Remedy