Configuration
34
en
Selecting or disabling the capture ranges (only applies for the measuring
mode):
The capture range is selected in the measuring mode (factory setting) and
permanently disabled in adjustment mode.
The CC-Vision PC software can be used to select or disable the capture ranges
for the measuring mode.
5.1 Instrument settings
The following parameters can be changed on an instrument:
5.2 Sensor settings
The following parameters can be changed on the sensors:
Name Range
Password Numerical range (3-digit)
Operating signal LED
1)
1) At least one of the two operating signals must be switched on.
Yes/No
Horn operating signal
1)
Yes/No
Switch-off mode "switch-off allowed" or
"switch-off prohibited" or
"switch-off prohibited at A2"
Shift length (TWA)
2)
2) Corresponds to the averaging time and is used to calculate the TWA exposure level.
60 - 14400 (in minutes)
(Setting for exposure alarm)
Short-term exposure level (STEL)
3)4)
3) Interpretation only if the sensor is designed for this.
4) Corresponds to the averaging time and is used to calculate the STEL exposure level.
0 - 15 (in minutes)
(setting for exposure alarm)
Name Range
Alarm threshold A1 (in measuring unit) Capture range – A2
Alarm threshold A2 (in measuring unit) A1 – measuring range end value
Interpretation type
1)
Inactive, TWA, STEL, TWA+STEL
Alarm threshold STEL
(in measuring unit)
1)
Capture range – full scale value