Dräger X-am 5600 - 6.8 Replacing the batteries; rechargeable batteries

Dräger X-am 5600
396 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
41
en
If functions in the quick menu are activated using the Dräger CC Vision PC
software, you can select these functions using the [ + ] key. If no functions
have been activated in the quick menu, the instrument remains in measuring
mode.
[OK] key to retrieve the retrieve the selected function.
Press the [ + ] key to cancel the active function and to switch to measuring
mode.
If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
6.8 Replacing the batteries / rechargeable batteries
!
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from
different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106
1)
(power one) or
Possible functions: 1. Bump test (configuration for bump test,
see technical manual)
2. Fresh gas adjustment
1)
1) The fresh-gas adjustment/zero-point adjustment is not supported by the CO2 channel of
the infrared sensor and the XXS O3. A zero-point adjustment of these sensors can be
made with the Dräger CC-Vision PC software. A suitable zero gas free from carbon
dioxide and ozone (e.g. N2) must be used.
3. Delete peak values
To display pump information,
see the Technical Manual
To activate or deactivate the pump,
see the Technical Manual
1) Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.