14
WARNINGS
DO NOT place pendant under or between objects. This may unintentionally press the buttons and may cause
injury or damage.
Electronic equipment may be inuenced by Radio Frequency Interference (RFI). Caution should be exer-
cised with regard to the use of portable communications equipment in the area around such equipment. If RFI
causes erratic behavior, unplug the electric bed IMMEDIATELY. Leave unplugged while transmission is in
progress.
Ensure all cables and cords are routed such that they will not become entangled or pinched. Otherwise dam-
age or injury may result.
If a button on the pendant sticks or does not release and/or one of the functions will not stop moving. Unplug
the power cord from the electrical supply outlet and take the bed out of service for repair.
If a liquid is spilled in or around the electric bed, unplug the electric bed before cleaning. Clean up the spill and
allow the electric bed or the area around the electric bed to dry thoroughly before using the electric controls
again.
Inspect the covering of the bed’s control panel and the patient control panel to assure that the covering is not
cracked or damaged.
Keep all electrical cords away from heated or hot surfaces.
Keep the cord away from heated surfaces.
NEVER operate if the unit has a damaged cord or plug. If it is not working properly, call a qualied technician
for examination and repair.
NEVER operate this bed if it has a damaged cord or plug, if it is not working properly, if it has been dropped,
damaged, or dropped into water. Return the bed to a service center for examination and repair.
On Full Electric beds, the Hi/Lo crank MUST be removed before the bed is used. Failure to remove the crank
may cause damage or personal injury.
Refer servicing to qualied personnel only. Grounding reliability depends upon a properly grounded wall outlet.
The pendant and power cords must be routed and secured properly to ensure that the cords DO NOT become
entangled, pinched and/or severed during operation of the electric bed.
Unplug from the outlet before servicing.
Unplug the power cord from its power source before performing any maintenance on the manual/electric bed.
DO NOT unplug the power cord from the junction box. Damage to cord will result.
Use caution when disconnecting the pendant. DO NOT press pendant buttons.