Installation, Operation and Service Manual LKP Series
© 2020 DynAmp, LLC Page 14
041736 N
Table 4.1 (A)
Measuring Head Resistance Chart
LKP-45
Measure Between Leads:
Desired Value
(Ohms)
Circuit
CABLE TO "A" HALF OF MEASURING HEAD
24 & 29 22-24 Coil Circuit - Channel #1A
25 & 30
10.5-12
Coil Circuit - Channel #2
26 & 32 10.5-12 Coil Circuit - Channel #3
27 & 31
22-24
Coil Circuit - Channel #4A
21 & 19 1000-1300 Hall Current - Channel # 1, 2, 3
11 & 12
240-550
Hall Signal - Channel #1
13 & 14 240-550 Hall Signal - Channel #2
15 & 16
240-550
Hall Signal - Channel #3
CABLE TO "B" HALF OF MEASURING HEAD
53 & 56 22-24 Coil Circuit - Channel #4B
52 & 58 10.5-12 Coil Circuit - Channel #5
49 & 55
10.5-12
Coil Circuit - Channel #6
48 & 54 22-24 Coil Circuit - Channel #1B
46 & 44
1000-1300
Hall Current - Channel # 4, 5, 6
40 & 41 240-550 Hall Signal - Channel #4
36 & 37
240-550
Hall Signal - Channel #5
38 & 39 240-550 Hall Signal - Channel #6