EasyManua.ls Logo

DynAmp LKP Series - Page 31

Default Icon
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation, Operation and Service Manual LKP Series
© 2020 DynAmp, LLC Page 15
041736 N
Table 4.1 (B)
Measuring Head Resistance Chart
LKP-60
Measure Between Leads:
Desired Value
(Ohms)
Circuit
CABLE TO "A" HALF OF MEASURING HEAD
24 & 29 22-24 Coil Circuit - Channel #1A
25 & 30
10.5-12
Coil Circuit - Channel #2
26 & 31 10.5-12 Coil Circuit - Channel #3
27 & 32
22-24
Coil Circuit - Channel #4A
21 & 19 1000-1300 Hall Current - Channel # 1, 2, 3
11 & 12
240-550
Hall Signal - Channel #1
13 & 14 240-550 Hall Signal - Channel #2
15 & 16
240-550
Hall Signal - Channel #3
CABLE TO "B" HALF OF MEASURING HEAD
53 & 56 22-24 Coil Circuit - Channel #4B
52 & 58 10.5-12 Coil Circuit - Channel #5
49 & 55 10.5-12 Coil Circuit - Channel #6
48 & 54 22-24 Coil Circuit - Channel #1B
46 & 44 1000-1300 Hall Current - Channel # 4, 5, 6
40 & 41 240-550 Hall Signal - Channel #4
36 & 37 240-550 Hall Signal - Channel #5
38 & 39 240-550 Hall Signal - Channel #6

Related product manuals