Rev. 01/11 - 2.12 - TFXL-X
2. On horizontal pipes, mount the transducer in an orientation such
that the cable exits at ±45 degrees from the side of the pipe. Do
not mount with the cable exiting on either the top or bottom of
the pipe. On vertical pipes the orientation does not matter.
3. Tighten the wing nuts so that the grease begins to flow out from
the edges of the transducer and from the gap between the
transducer halves. Do not over tighten.
4. If Signal Strength is less than 7, remount the transducer at
another location on the piping system.
5. If Signal Strength is greater than 100, contact Dynasonics to
obtain a lower power strategy to load into the TFXL meter.
NOTE: If a DTTS/DTTC small pipe transducer was purchased
separately from the TFX meter, the following configuration
procedure is required.
DTTS/DTTC Small Pipe Transducer Configuration Procedure
1. Establish communications with the transit time meter. See Part 3.
2. From the Tool Bar select Calibration. See Figure 2.7.
PART 2 - TRANSDUCER INSTALLATION
Figure 2.7
Data Display Screen