Rev. 01/11 - 2.13 - TFXL-X
3. On the pop-up screen, click Next button twice to get to Page 3 of 3.
See Figure 2.8.
4. Click Edit.
5. If calibration point is displayed in Calibration Points Editor
screen, record the information, highlight and click Remove. See
Figure 2.9.
6. Click ADD…
7. Enter Delta T, Uncalibrated Flow, and Calibrated Flow values
from the DTTS/DTTC calibration label, the click OK. See Figure
2.10.
8. Click OK in the Edit Calibration Points screen.
9. Process will return to Page 3 of 3. Click Finish. See Figure 2.8.
PART 2 - TRANSDUCER INSTALLATION
Figure 2.8
Calibration Page 3 of 3
Figure 2.9
Calibration Points Editor