Rev. 01/11 - 2.14 - TFXL-X
PART 2 - TRANSDUCER INSTALLATION
10. After “Writing Configuration File” is complete, turn power off.
Turn on again to activate new settings.
Mounting Transducers in Z-Mount Configuration
Installation on larger pipes requires careful measurements to the
linear and radial placement of the DTTN and DTTH transducers.
Failure to properly orient and place the transducers on the pipe may
lead to weak signal strength and/or inaccurate readings. The section
below details a method for properly locating the transducers on
larger pipes. This method requires a roll of paper such as freezer
paper or wrapping paper, masking tape and a marking device.
1. Wrap the paper around the pipe in the manner shown in Figure
2.11. Align the paper ends to within ¼ inch (6mm).
2. Mark the intersection of the two ends of the paper to indicate the
circumference. Remove the template and spread it out on a flat
surface. Fold the template in half, bisecting the circumference.
See Figure 2.12 on page 2.15.
3. Crease the paper at the fold line. Mark the crease. Place a mark
on the pipe where one of the transducers will be located. See
Figure 2.1 on page 2.8 for acceptable radial orientations. Wrap
DTTN Z-Mount
Transducer
Installation
Figure 2.10
Edit Calibration
Points
Figure 2.11
Paper Template Alignment

Other manuals for dynasonics TFXL series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the dynasonics TFXL series and is the answer not in the manual?

dynasonics TFXL series Specifications

General IconGeneral
Branddynasonics
ModelTFXL series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals