2.3 SYSTEM COMPONENTS
2.2 FEED WATER REQUIREMENTS
Parameter Value
1
pH 6.5…8.5
TDS, mg/L (ppm) <1500
Total hardness, °dH
ppm CaCO
3
<30
<500
Active chlorine, mg/L (ppm) <0.5
Total iron, mg/L (ppm) <0.3
Total manganese, mg/L (ppm) <0.1
COD, mg O
2
/L (ppm O
2
) <5
Total microbial count (TMC), CFU per 1 mL <100
E. coli, CFU per 100 mL None
1
If feed water quality does not meet specied requirements, membrane and lter service life may decrease. If
aiming to purify raw water (well water/ground water etc.), it is recommended to carry out detailed water anal-
ysis. If any constuents exceed the above limitaons, it is advisable to install a special lter upstream of your
reverse osmosis system. Consult professional water treatment specialist for proper selecon of an appropriate
lter for your water.