EasyManua.ls Logo

ECS Electronics AdvancedIQ AG-6500 - Operating Instructions; Calibration

ECS Electronics AdvancedIQ AG-6500
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Operating Instructions
1. Once unit is verified to be wired correctly, power unit on. The status LED light will repeatedly flash
green two times quickly for two (2) minutes. The sensor heater is warming up during this period.
2. After the two-minute warm-up period, the sensor and status LED light will flash green one time
repeatedly indicating normal operation. At this time, the display will show the current average
concentration of the feed gas.
Calibration
NOTE: Please contact ECS before any calibration adjustment
1. Allow SMART Gas Analyzer to sample fresh compressed air for a minimum of ten (10) minutes.
2. Switch N
2
/O
2
switch (SW 1-1) to O
2
position.
3. If digital display not reading approximately 20.9%, switch CAL switch (SW 2-4) to CAL position.
4. Allow SMART Gas Analyzer to remain in Calibration Mode and sample the gas for a minimum of
ten (10) minutes to allow SMART Gas Analyzer to recalibrate.
5. Switch CAL switch (SW 2-4) to Off position.
6. Verify digital display is reading approximately 20.9%.
7. Switch N
2
/O
2
switch (SW 1-1) to N
2
position.
PC Board - LED Locations
Alarm LED
Over LED
SW1
SW2
Tx LED
Rx LED
5V Pwr
LED
Heater
LED
Status
LED
Display
Port
Zr02
Port