CADENCE II Fetal Monitor User Manual
- 65 -
To replace a fuse:
A Place the monitor on a flat surface.
A With a flat-head screw driver, push the fuse in for about 1 mm and then unscrew it
anticlockwise.
A Remove the old fuse and replace it with a new fuse that is supplied by the manufacturer or of
the same specifications.
A Push the new fuse into the socket for about 1 mm and then screw it clockwise back in
position.