CADENCE II Fetal Monitor User Manual
- 40 -
7.5.4 Switching the UA Probe Detecting Alarm On or Off
The monitor detects if the TOCO transducer is well connected. When the transducer is unplugged,
the monitor gives an alarm. You can switch this alarm on or off.
1 Press the SETUP key twice to open the Alarm Configuration menu.
2 Press the UP key or DOWN key to move the cursor to UAProbe; press the CHANNEL
key.
3 Press the UP key or DOWN key to toggle between On and Off (default).
4 Press the CHANNEL key to confirm selection.
5 Press the AUTO key to exit.
7.6 Monitoring Uterine Activity Internally (optional)
7.6.1 Parts Required
1) Disposable intrauterine pressure catheter ACCU-TRACE
IUPC (“IUPC” for short)
2) Reusable intrauterine pressure connecting cable (“connecting cable” for short)
3) Reusable intrauterine pressure cable (“IUP cable” for short)
7.6.2 Directions for Use of IUPC
Preparation
1) Gather supplies: ACCU-TRACE IUPC, reusable cable, and amnioinfusion supplies if
needed.
2) Open the sterile ACCU-TRACE IUPC package.
Insertion
NOTE: This product is designed for use with the introducer.
3) Using aseptic technique, remove the catheter from the package.
4) Perform vaginal exam to ensure ruptured membranes and adequate dilation.
5) Advance the catheter tip to the cervical os along the examination hand, using the hand as a
guide. Do not advance the introducer through the cervix.
6) Continue to gently advance the catheter tip through the cervical os and feed the catheter into
the intra-amniotic cavity until the 45cm mark is at the introitus. If the 45cm mark is not
clearly visible, stop advancing when the
symbol on the catheter meets the introducer.
NOTE: For easier insertion, do not twist the catheter in the introducer.
7) The IUPC may spontaneously fill with amniotic fluid. This can be seen in the clear lumen of
the catheter. The filter cap will prevent the amniotic fluid from leaking.
8) Slide the introducer out of the vagina along the catheter. When the introducer is completely
out of the vagina, slide thumb between catheter and introducer tab, which will begin to
separate the introducer from the catheter. (See figure 7-3)