About this Manual
P/N: 01.54.455473-10
Release Date: Jan. 2012
© Copyright EDAN INSTRUMENTS, INC. 2012. All rights reserved.
Statement
This manual will help you understand the operation and maintenance of the product better. It is
reminded that the product shall be used strictly complying with this manual. Users operation
failing to comply with this manual may result in malfunction or accident for which EDAN
INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information are contained in this manual, the user shall
not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
Responsibility of the Manufacturer
EDAN only considers itself responsible for any effect on safety, reliability and performance of
the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by EDAN, and
The electrical installation of the relevant room complies with national standards, and
The instrument is used in accordance with the instructions for use.
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other
information to help qualified technician to maintain and repair some parts, which EDAN may
define as user serviceable.
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce

Table of Contents

Other manuals for EDAN iM80

Questions and Answers

  • R
    Robert ColemanSep 7, 2025
    Why is there no CO waveform on my EDAN Medical Equipment?
    • L
      Leslie ReynoldsSep 7, 2025
      If there is no CO waveform on your EDAN Medical Equipment, there may be a bad connection of the CO module. Power off the device and reconnect the CO module. If the problem persists, the CO module may be defective and require replacement.
  • J
    John GouldAug 25, 2025
    What to do if EDAN iM80 Monitor displays 'PULSATILE PRESSURE ZERO FAIL'?
    • K
      Keith RamosAug 25, 2025
      If the EDAN Monitor displays 'PULSATILE PRESSURE ZERO FAIL,' ensure the transducer is vented to air and not connected to a patient. Then, try again.
  • S
    sunderwoodAug 25, 2025
    What to do if EDAN iM80 displays 'Art ZERO FAIL'?
    • S
      Sabrina LeeAug 26, 2025
      If the EDAN Monitor displays 'Art ZERO FAIL', make sure that the transducer is not attached to the patient.
  • K
    keithmerrittSep 16, 2025
    Why does the AG measurement numeric value have an error on EDAN Medical Equipment?
    • L
      Laurie HerreraSep 16, 2025
      If the AG measurement numeric value has an error on your EDAN Medical Equipment, the GAS module may not have been zeroed for a long time, resulting in inaccurate measurements. Zero the GAS module. Also, ensure the system is in measurement mode and allow a 10-minute warm-up before measuring.
  • J
    Jacob PerezSep 23, 2025
    What to do if RESP waveform amplitude is faint on EDAN Medical Equipment?
    • L
      lawrencemariaSep 23, 2025
      Adjust waveform amplitude multiple in RESP menu.
  • K
    Kaitlin BarrSep 12, 2025
    Why does paper advance lopsidedly on EDAN Medical Equipment?
    • M
      Manuel LarsonSep 12, 2025
      If the paper advances lopsidedly in your EDAN Medical Equipment, the assembly or location of the recorder may be incorrect. Calibrate the assembly of the recorder.
  • N
    nberrySep 2, 2025
    Why is there no NIBP numeric value on my EDAN Medical Equipment?
    • K
      Kathryn WolfeSep 3, 2025
      If there's no NIBP (Non-Invasive Blood Pressure) numeric value displayed on your EDAN Medical Equipment, it could be due to a loose cuff or patient movement. Make sure the cuff is properly secured and ask the patient to remain still.
  • K
    Katie TaylorSep 3, 2025
    What causes fuses to blow after switching on EDAN Medical Equipment?
    • B
      Brad GomezSep 3, 2025
      If the fuses are blown after switching on your EDAN Medical Equipment, further examination after switching on is required to identify any short circuits in the power or other parts.
  • N
    Nicole DukeSep 23, 2025
    What to do with frequent patient activity with EDAN Medical Equipment?
    • H
      holly25Sep 23, 2025
      Quiet the patient.
  • H
    Heather Garcia DVMAug 29, 2025
    What to do if EDAN Medical Equipment LCD is black, power is on, and the fan runs?
    • E
      elizabeth58Aug 29, 2025
      If your EDAN Medical Equipment's LCD has no display but the power indicator is lit and the fan runs, check if the keys sound properly. If they do, the LCD might be defective. However, if the keys don’t sound, the main control board might be the problem, and you should replace it.

Summary

Chapter 1 Indication for Use and Safety Guidance

1.1 Indication for Use

Describes the intended medical applications of the monitor.

1.2 Safety Guidance

Provides critical safety warnings and precautions for operation.

Chapter 4 Alarms

4.1 Alarm Category

Classifies alarms into physiological, technical, and prompts.

4.2 Alarm Levels

Categorizes alarms by severity: high, medium, low.

4.3 Controlling Alarm

How to manage alarms, including mute and switching off.

Chapter 5 Alarm Information

5.1 Physiological Alarm Information

Details physiological alarms, their causes, and levels.

5.2 Technical Alarm Information

Details technical alarms, their causes, and actions.

Chapter 6 Managing Patients

6.1 Admitting a Patient

Steps for entering patient information for monitoring.

Chapter 8 Monitoring ECG

8.2 ECG Safety Information

Crucial safety precautions for ECG monitoring.

Chapter 9 Monitoring RESP

9.2 RESP Safety Information

Safety guidelines specific to respiration monitoring.

9.4 Electrode Placement for Monitoring Resp

Correct placement of electrodes for respiration monitoring.

Chapter 10 Monitoring SpO2

10.2 SpO2 Safety Information

Important safety precautions for SpO2 monitoring.

Chapter 11 Monitoring PR

11.5 Selecting the Active Alarm Source

How the monitor chooses between ECG/Pulse for alarms.

Chapter 12 Monitoring NIBP

12.2 NIBP Safety Information

Critical safety warnings for NIBP measurements.

12.6 Measurement Procedures

Detailed steps for performing NIBP measurements.

Chapter 13 Monitoring TEMP

13.2 TEMP Safety Information

Safety precautions for temperature monitoring.

Chapter 14 Monitoring Quick TEMP

14.2 Quick TEMP Safety Information

Safety guidelines for quick temperature measurements.

Chapter 15 Monitoring IBP

15.2 IBP Safety Information

Safety precautions for IBP monitoring.

Chapter 16 Monitoring CO2

16.2 CO2 Safety Information

Safety guidelines specific to CO2 monitoring.

Chapter 17 Monitoring C.O.

17.2 C.O. Safety Information

Safety precautions for Cardiac Output monitoring.

Chapter 18 Monitoring AG

18.2 Safety Information

Safety guidelines for AG monitoring modules.

Chapter 20 Review

Chapter 21 Calculation and Titration Table

21.1 Drug Calculation

Performing drug dosage calculations.

21.1.1 Calculation Procedures

Steps for executing drug calculations.

Chapter 25 Care and Cleaning

25.1 General Points

General precautions for cleaning the monitor and accessories.

25.2 Cleaning

Detailed instructions for cleaning the monitor enclosure and screen.

25.2.1 Cleaning the Monitor

Specific cleaning agents and procedures for the monitor.

Chapter 26 Maintenance

26.1 Inspecting

Items to check during general monitor inspection.

Chapter 28 Accessories

C Default Settings

EDAN iM80 Specifications

General IconGeneral
Display TypeTFT LCD
RespirationYes
SpO2Yes
NIBPYes
TemperatureYes
Pulse RateYes
EtCO2Optional
BatteryRechargeable lithium-ion battery
ECG3/5 lead ECG
Optional ParametersCardiac Output
NetworkingWi-Fi, Ethernet
Weight2.5 kg