Patient Monitor User Manual Product Specification
- 155 -
Cuff pressure measuring range
Maximum standard deviation
30s ~ 45s (depend on HR/motion disturbance)
Overpressure protection (Dual overpressure protection)
±3bpm or 3.5%, whichever is greater
A.10.2 NIBP from M3600 Module
Measuring Interval in AUTO Mode
1/2/3/4/5/10/15/30/60/90 min, 2/4/8h
Adult/ Pediatric mode: 40bpm ~ 200bpm
Neonatal mode: 40 bpm ~ 240bpm
± 2 bpm or 2% of the readings
DIA: 40 mmHg ~ 200 mmHg
MAP: 45 mmHg ~ 235 mmHg
DIA: 20 mmHg ~ 90 mmHg
MAP: 30 mmHg ~ 100 mmHg