Patient Monitor User Manual Product Specification
- 149 -
ECG Analog Output
Bandwidth (-3dB)
Diagnosis: 0.05Hz ~ 100Hz
Monitor: 0.5Hz ~ 40Hz
Surgery: 1Hz ~ 20Hz
Maximum transmission delay
500ms (in diagnostic mode, and with notch off)
Sensitivity
1V/mV ±10%
PACE rejection/enhancement
Without Pace enhancement or pace rejection
Defib Sync Pulse
Output wave
Square pulse
Output impedance
50Ω
Delay from R-wave peak to start
of pulse
35ms
Amplitude
High level: 3.5 to 5 V, providing a maximum of 1 mA
output current;
Low level: < 0.5V, receiving a maximum of 5 mA input
current.
Minimum required R wave
amplitude
0.3mV
Pulse width
100ms ± 10%
Limited current
15 mA rating
Rising and falling time
< 1 ms
A.8.2 12-lead Monitoring
Lead Mode
3-Lead: I, II, III
5-Leads: I, II, III, aVR, aVL, aVF, V
12-Lead: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform
3-Lead: 1-channel waveform
5-Lead: 2-channel waveform, a maximum of seven waveforms;
12-Lead: 2-channel waveform, a maximum of 13 waveforms.
Lead naming style
AHA, IEC
Display Sensitivity
1.25mm/mV (×0.125),
2.5mm/mV (×0.25), 5mm/mV (×0.5),
10mm/mV (×1), 20mm/mV (×2), AUTO gain
Bandwidth (-3dB) Diagnosis: 0.05Hz to 150Hz