Patient Monitor User Manual Default Settings
- 176 -
Anes Agent
0%
Alarm High Limit (EtCO
2
)
50
50
45
Alarm Low Limit (EtCO
2
)
15
20
30
Alarm High Limit (FiCO
2
)
4
4
4
Alarm High Limit (AWRR)
30
30
100
Alarm Low Limit (AWRR)
8
8
30
Sweep
12.5mm/s
Amplitude
Low
C.12 C.O.
C.O. Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Alarm High Limit (TB)
43.0
43.0
43.0
Alarm Low Limit (TB)
23.0
23.0
23.0
Injectate Temperature
Source
Auto
Temperature Unit
°C
Interval
30
Constant
0.542
C.13 AG
AG Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Work Mode
Measure
Apnea Time
20s
Unit
%
O
2
Compensate
OFF
Anes Agent
HAL
Alarm High Limit (EtAA)
8.0
8.0
8.0