Patient Monitor User Manual Monitoring C.O.
- 101 -
Thermodilution Catheter Site
17.8 Setting the Computation Constant
The computation constant is associated with catheter and injectate volume. When the catheter is
changed, please adjust Constant in the C.O. Settings menu based on product description
provided by the manufacturer.
17.9 Recording C.O. Measurements
C.O. measurement can be recorded by the recorder. To record the C.O. measurement, please
select Record in the C.O. Measure menu.
17.10 Setting INJ. TEMPSource
To change the INJ Temp Source, please:
1 Select Inj Temp Source in the C.O. Settings menu;
2 Select Auto or Manual from the list;
Manual: directly displaying the injectate temperature from INJ. TEMP.
Auto: indicating the system obtains the injectate temperature through sampling.
Right ventricle