Patient Monitor User Manual Alarm Information
- 41 -
5.3 Prompts
Message Cause
ECG Arr Learning
The QRS template building required for Arr. Analysis is in
process.
SpO
2
Search Pulse
SpO
2
module is analyzing the patient signal and searching
for the pulse to compute the saturation, when sensor is
connected with patient.
Manual Measuring In manual measuring mode
Continual Measuring In continuous measuring mode
Auto Measuring In automatic measuring mode
Measure Abort Measurement over
Calibrating During calibrating
Calibrate Abort Calibration over
Leak Testing During pneumatic test
Leak Test Ok NIBP module has passed leak test.
Leak Test Abort Pneumatic test over
Resetting NIBP module in resetting
Please Start NIBP module is in idle status.
Done NIBP measurement successfully done
Continual Measuring NIBP module performs continual measuring
Stat Measuring NIBP module performs STAT function.
Please Switch To Maintain
Mode
NIBP module is in normal mode, the user can’t start leak
test and pressure calibration. Please enter User
Maintain > NIBP Maintain and switch to Maintain
Mode to perform leak tset or pressure calibration.
Please Switch To Normal Mode
NIBP module is in maintain mode, the user can
t start
blood pressure measurement. Please enter User
Maintain > NIBP Maintain and switch to Normal Mode
to perform blood pressure measurement.
Quick TEMP Is Warming Up Quick TEMP Is Warming Up.
Place Probe On Measure Place Probe isn’t placed on the measurement site.
CO
2
Standby
Turn from measuring mode to standby mode, making the
module in energy-saving status.
CO
2
Sensor Warm Up The CO
2
module is in warm-up state.
Excessive C.O. Temp TB measuring value is beyond measuring range.