Patient Monitor User Manual Alarm Information
- 40 -
Probe data error
Offline: NTC resistance >R
0 °C; Short: NTC resistance
<R+100 °C.
Med
Put the sensor into
the sensor bracket,
take it out and
measure again. If the
problem persists, stop
using measuring
function of TEMP
module, and notify
biomedical engineer
or manufacturer’s
service staff.
Probe heater error Single failure Med
Put the sensor into
the s
ensor bracket,
take it out and
measure again. If the
problem persists, stop
using measuring
function of TEMP
module, or notify
biomedical engineer
or manufacturer’s
service staff.
Probe temp too high
The original temperature of
sensor >+33 °C & ≤+40 °C.
Low
Put the sensor into
the sensor bracket,
measure again after
the sensor
temperature reaches
normal value.
Temp Sensor Off
After the sensor temperature
reaches Predict value, it
descends to the value lower
than Predict value.
Med
Reconnect the sensor
and make sure that
the cable is properly
connected.
Battery Low
Battery Low
High
Please change the
battery or charging.
Recorder Out Of Paper
Recorder Out Of Paper Low
Please install the
paper
Recorder setup needed
The user presses the
RECORD button when the
monitor is not installed with
a recorder.
Low
Notify the
manufacturer’s
service staff to install
and set the recorder.