Patient Monitor User Manual Product Specification
- 163 -
reading
3)
DES
4)
15 vol%
+12% of
reading
3)
_ 1)
_ 1)
_ 1)
Xe(Xenon)
4)
80 vol%
-10% of
reading
3)
_ 1)
_ 1)
He(Helium)
4)
50 vol%
-6% of
reading
3)
_ 1)
_ 1)
Metered dose
inhaler
propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4)
0.3 vol%
_ 1)
_ 1)
_ 1)
_ 1)
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CH
3
COCH
3
(Acetone)
4)
1 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CH
4
(Methane)
4)
3 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CO(Carbon
monoxide)
5)
1 vol%
_ 1)
_ 1)
_ 1)
_ 1)
NO(Nitrogen
monoxide)
0.02 vol%
_ 1)
_ 1)
_ 1)
_ 1)
O
2
5)
100 vol%
_ 2)
_ 2)
_ 1)
_ 1)
Note 1: Negligible interference, effect included in the specification Accuracy, all conditions
above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included in the
specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the
CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0 vol% CO
2
and 50
vol% Helium, the actual measured CO
2
concentration will typically be (1-0.06)*5.0 vol% =4.7
vol% CO
2
.
Note 4: According to the EN ISO 21647 standard.
Note 5: In addition to the EN ISO 21647 standard.
A.17.2 Phasein Mainstream
Module Type
IRMA AX+
Displaying the concentration of CO
2
, N
2
O
and two anaesthesia agent and indentifying
two anaesthesia agent
Measurement
CO
2
, N
2
O, O
2
, HAL, Isoflurane(ISO), Enflurane(ENF),

Table of Contents