SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 45 -
1. System Troubles
Conditions
Possible Causes
Actions
When the mains supply is
used, you can not turn on
the electrocardiograph.
Fuse failure
Power board or main
board failure.
Silicone keyboard
failure
Replace the fuse
Replace the power board
or the main board
Replace the silicone
keyboard
When the battery is used,
you can not turn on the
electrocardiograph.
Battery failure
Main board or power
board failure
Silicone keyboard
failure
Replace the battery
Replace the main board or
the power board
Replace the silicone
keyboard
The electrocardiograph fails
to function.
Strong interference of
electric networks
Main board failure
The plug-in unit on
the main board or the
power supply failure
Inspect the power supply
and grounded system.
Replace the main board
Replace or repair the
plug-in unit
Fuse is burned when you
turn on the
electrocardiograph.
The power supply or
other components
short-circuit.
Open the
electrocardiograph for
further examination
Fuse is burned when the
device connects to a certain
component.
This component
short-circuits.
Replace this component.
2. Display Troubles
Conditions
Possible Causes
Actions
Some characters displayed
on the LCD screen are
missing.
LCD screen failure.
Replace the LCD screen
The LCD screen is dark.
The brightness of the
LCD screen is not
adjusted well.
LCD screen failure
Adjust the brightness of the
LCD screen in the Display
Setup window.
Replace the LCD screen