SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 46 -
3. Transmission Troubles
Conditions
Possible Causes
Actions
Fail to transmit ECG data
through the net port
Fail to copy data from the
electrocardiograph to the U
disk
Setup error
Connection cable
failure
Port failure
U disk failure
Examine whether the
transmission settings are
correct. (The first three
sections of Local IP and
Remote IP must be set the
same as the first three
sections of the IP of the
PC.)
Replace the connection
cable
Replace the port
Replace the U disk
4. Operation Troubles
Conditions
Possible Causes
Actions
Key failure
The key board is
damaged
Main board failure
Replace the key board
Replace the main board
No key beep or key beep is
raucous
Speaker failure
Replace the speaker
The electrocardiograph can
not print reports.
The recorder paper
runs out.
Print head failure
Main board failure
Recorder related cable
failure
Install the recorder paper
Replace the print head
Replace the main board
Replace or repair the
recorder related cable