SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 46 -
3. Transmission Troubles
Fail to transmit ECG data
through the net port
Fail to copy data from the
electrocardiograph to the U
disk
① Setup error
② Connection cable
failure
③ Port failure
④ U disk failure
① Examine whether the
transmission settings are
correct. (The first three
sections of Local IP and
Remote IP must be set the
same as the first three
sections of the IP of the
PC.)
② Replace the connection
cable
③ Replace the port
④ Replace the U disk
4. Operation Troubles
① The key board is
damaged
② Main board failure
① Replace the key board
② Replace the main board
No key beep or key beep is
raucous
The electrocardiograph can
not print reports.
① The recorder paper
runs out.
② Print head failure
③ Main board failure
④ Recorder related cable
failure
① Install the recorder paper
② Replace the print head
③ Replace the main board
④ Replace or repair the
recorder related cable