SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 47 -
The printed report is illegible
The rubber roller of
the recorder casing
has stain.
The elasticity of the
spring of the recorder
frame is not good.
Clean the rubber roller
Replace the spring
Some characters or
waveforms printed are
missing.
The rubber roller of
the recorder casing
has a stain.
Print head failure.
Replace the rubber roller
Replace the print head
Fail to detect paper
There is a stain on the
detecting position of
the printing bracket.
Reflective
photosensor failure
Main board failure
Clean the stain on the
detecting position of the
printing bracket
Replace the reflective
photosensor
Replace the main board
5. Parameter Troubles
Conditions
Possible Causes
Actions
No ECG waveform
Defective connection
between the patient and
the electrodes
Defective connection
between the patient
cable and the unit
ECG board failure
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Connect the patient cable
to the unit
Replace the ECG board
ECG waveform is abnormal
or disturbed
Defective connection
between the patient and
the electrodes
Patient cable failure
ECG board failure
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Replace the patient cable
Replace the ECG board
Heart rate is not accurate
Waveform measuring
failure
Adjust connections