SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 48 -
ECG waveform has burrs
Defective connection
between the patient and
the electrodes
ECG board failure
Electrical interference
from another device
(microwave oven,
cellular phone, wireless
device, etc.)
An
improperly-grounded
electrical device near
the electrocardiograph
Attach the electrodes to
the patient again or clean
the electrode area on body
surface with alcohol
Replace the ECG board
Look for devices that
could be causing
electrical interference,
and then unplug the
devices. Or run the
electrocardiograph on the
battery power.
Lay the lead wires
alongside the limbs and
away from any electrical
devices. Turn on the AC
filter on the Filter Setup
window