ABC Analog Hold
Typing the letters ‘AH’ will toggle the ABC Hold feature on or off. If ABC Hold is
on, the analog and serial outputs will be held at the values just prior to initiating
the ABC Cycle. The serial output will contain the message “ABC Hold”. The hold
will be released when the ABC Cycle is complete and the instrument has
stabilized back on the dew point.
Date
Type the letter “D” and then press ENTER.
The DATE format is: MONTH/DAY/YEAR
“01/01/03” = January 1, 2003
“12/31/02” = December 31, 2002
NOTE: Two digits must be used for each entry field and separated by a
backslash character.
Output Interval
This is the interval in seconds between automatic data output transmissions of
the serial data output. The time range is from 0 to 3600 seconds.
Type the letter “O” and press the ENTER key. Enter the desired interval in
seconds and press ENTER.
Poll for Output
This command requests the RH CALIBRATOR to send serial data at any time
and is independent of the automatic interval.
Type the letter “P” and press the “ENTER” key.
Manual Heat
This command toggles the Sensor’s heater on or off. It can be initiated at any
time and can be used to clear excessive moisture from the mirror in flooding
situations.
Type the letters “MH” and press the ENTER key. The RH CALIBRATOR will
display a flashing MAN MAX HEAT and the temperature will rise. To turn off the
MANUAL HEAT mode, enter “MH” again.
Manual Cool
This command toggles the Sensor’s cooler on or off. It can be used to test the
maximum depression of the Sensor.
Type the letters “MC” and press the ENTER key. The RH CALIBRATOR will
display a flashing MAN MAX COOL and the temperature will decrease. To turn
off the MANUAL COOL mode, enter “MC” again.