29
NOTE: After an extended period of time in the cool mode, excess moisture or
frost will form on the mirror. It may be necessary to “Manually Heat” the mirror to
shorten the drying time.
Time
Type the letter “T” and press ENTER.
The time format is: Hours:Minutes:Seconds. Hours is expressed in 24-hour
military time.
Enter the desired time with colon delimiters:
Examples:
“00:00:00” = 12 midnight
“23:59:59” = 11:59:59 pm
Units (F,C)
The RH CALIBRATOR can display temperature in either Degrees C or Degrees
F. The display and RS-232 data will reflect the selection.
Type the letter “U” and press ENTER. Type “C” or “F” and press ENTER.
Status Report
This command gives the user a report of all of the current settings. To get a
status report, type the letters “ST” and press the ENTER key.