Pump start / stop and control
To St
art or Stop the pump the PDT must be in control, shown by the Local Control LED
being illuminated.
To take or release control briey press Control.
If something else is in control, error message 'Control locked' appears. Refer to 'Control
Holder' in the Status menu.
Press Start
buon PDT displays:
Table 18
Pump start control
Pump State
Local Control
LED
PDT display Operator Pump response
Stopped On START MENU
Start Pump
Press Enter to
conrm
Pump Starts
Running On Pump Running
Press CANCEL
Press CANCEL No change (run-
ning)
Stopped or
Running
O No PDT Control
Press CANCEL
Press CANCEL No change
Press Stop
buon PDT displays:
Table 19
Pump stop and control
Pump state
Local con-
trol LED
PDT display Operator
Pump re-
sponse
Stopped On
PUMP Stopped
Press CANCEL
Press CANCEL
No change
(stopped)
Running On
STOP MENU Fast
Shutdown
If Auto Shutdown is
necessary press down
arrow. Press ENTER to
conrm
Pump stops
Stopped or Run-
ning
O No PDT Control
Press CANCEL
Press CANCEL No change
Warning and alarm display and acknowledgement
Each new warning and alarm is displayed as it occurs, overwring any current display
text unless another unacknowledged warning or alarm is being displayed. The
corresponding Warning and Alarm LED ashes to indicate a new warning and alarm.
Press Enter to acknowledge the warning and alarm currently displayed. The Warning and
Alarm LED stops ashing. If available the display will show the suggested acon. Press
Enter again to clear.
If there is another new warning or alarm, the Warning and Alarm LED will connue to
ash and this is then displayed. Otherwise the display will revert to the original text from
before the warning(s) and alarm(s) occurred.
If there are warnings and alarms sll present, but all acknowledged, the corresponding
LED remains connuously illuminated. The text indicang acknowledged warning and
10/2020 - ©Edwards Limited