Table 3
Checklist of components
Quanty
Descripon
Check (ü)
1
Vibraon isolator B58053400
r
1
Ins
trucon manual
r
5.3 Install the Vibr
aon Isolat
or
WARNING: HEAVY OBJECT
Risk of muscle strain or back injur
y. The mass of the Vibraon Isolator is 10.7 kg.
WARNING: HAZARDOUS VOLTAGE
Risk of injury or damage t
o equipment. Make sure that you switch o the pump and
isolate the controller from the power supply.
Refer to the instrucon manual supplied with your nEXT pump to determine the
minimum number of clamps required to t the vibraon isolator to the pump inlet
ange. You must use the same number of clamps to t the upper ange of the vibraon
isolator to your vacuum system.
To install the vibraon isolator, do the steps that follows:
1. Switch o the pump.
2. Isolate the pump controller from the electrical supply and wait unl the pump has
stopped r
otang.
3. Install the vibraon isolator between the inlet ange of the nEXT pump and the
vacuum system.
4. Install a exible below between the nEXT pump backing port.
Note:
The vibraon isolator is designed to operat
e in vercal direcon. If you install the
vibraon isolator in any other direcon, the funconal eciency will be eected.
Refer to Figure: Installaon orientaon.
09/2021 - ©Edwards Limited
Page 12B80000860_A
B80000860_A - Installaon