EasyManuals Logo
Home>Edwards>Water Pump>nXDS Series

Edwards nXDS Series User Manual

Edwards nXDS Series
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
© Edwards Limited 2017. All rights reserved. Page 35
Edwards and the Edwards logo are trademarks of Edwards Limited.
Maintenance
A735-01-880 Issue C
Table 15 - Flashing error codes
Error flash
position
Error flash
sequence
Comment Actions
0 ssssss Overload timeout. Check whether the pump is not
running under constant high pressure
or the inlet or outlet is not blocked.
1 Lsssss Controller software error. Cycle the power to the pump and see
whether the error code appears again.
If it does, contact the supplier or
Edwards.
2 sLssss Controller failed internal
configuration and calibration
operation.
Cycle the power to the pump and see
whether the error code appears again.
If it does, contact the supplier or
Edwards.
3 ssLsss Acceleration timeout. Check whether the pump is not
running under constant high pressure
or the inlet or outlet is not blocked.
4 sssLss Over-current trip activated, or other
hardware fault.
Cycle the power to the pump and see
whether the error code appears again.
If it does, contact the supplier or
Edwards.
5 ssssLs Self test fault. Cycle the power to the pump and see
whether the error code appears again.
If it does, contact the supplier or
Edwards.
6 sssssL Serial Control Mode Interlock. Re-activate serial enable and send a
serial command to clear the error
code.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nXDS Series and is the answer not in the manual?

Edwards nXDS Series Specifications

General IconGeneral
BrandEdwards
ModelnXDS Series
CategoryWater Pump
LanguageEnglish

Related product manuals