9.4 Operation steps
1. press +/-to choose a memory number
2.Press the S key to select a function that needs to be set
3. Press +/- to set the parameter that user need.
4. Every time the parameters are changed, it will be saved automatically.
*If exceed 5 seconds without any operation ( factory setting is 5 secondsit could
be change),display will be switched to standby interface.
Speed setting
value:120, 150, 200, 250, 300, 350, 400, 450, 500, 550, 600,
650,700,800,900,1000(rpm)
Reverse torque value setting
value:0.5, 0.6,0.8, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0 (N·cm)
Direction setting
FwdClockwise rotation RevCounter clockwise rotation
Note
Please set the parameters according to the file manufacturer's
recommendations.
Use of torque reversal function can effectively protect the file from separating
within the canal.
If the torque reversal is too frequent during the use, please recapitulate, irrigate
and lubricate the root canals or increase the torque as per file manufacturer’s
recommendations.