37
Immunity
test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Conducted
dis-turbances
induced by RF
fields
IEC 61000-4-
6
3 V
0.15 MHz 80
MHz, 6 V in
ISM bands be-
tween 0.15
MHz and 80
MHz, 80 % AM
at 1 kHz
3 V/m, 80 MHz
2,7 GHz,
80 % AM at 1
kHz
See the RF
wireless
communication
equipment
table in
"Recommende
d minimum
separation
distances"
3 V
Portable and mobile RF
communications equipment
should be usedno closer to any
part of the E-connect , including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended minimum
separation distances
See the RF wireless
communication equipment table in
"Recommended minimum
separation distances"
Radiated RF
EM fields
IEC 61000-4-
3
3V/m
Proximity
fields from RF
wireless
communicatio
n equipment
Complies
IEC 61000-4-
3
1. Use of accessories and cables other than those specified or provided by the
manufacturer of E-connect could result in increased electromagnetic emissions
or decreased electromagnetic immunity of E-connect and result in improper
operation.