8 Maintenance
Page 50 / 66
z Do not use bleach or chloride disinfectant materials.
8.3 Disinfection components
Disinfection components
Sensor
USB Extension Cable
Sensor Mounting
Bracket
WARNING
Before first use, after each use and whenever there is a risk of contamination,
disinfect the above components.
Reprocessing Instructions
Preparation before cleaning:
Immediately after using, remove the used disposable sleeve. Disconnect all
plug connections. Put the sensor and USB Extension Cable in container for
transportation.
WARNING
Make sure that used sleeves are disposed of as infected waste which is
potentially biologically hazardous.
Transportation:
Safe storage and transportation to the reprocessing area to avoid any damage
and contamination to the environment.
Cleaning and Disinfection:
1. Clean the sensor, USB extension cable and mounting bracket surface with a
cloth moistened in mild soapy water. After this, dry the components completely
with a clean lint-free cloth.