8 Maintenance
Page 52 / 66
1. Only the above components can be autoclaved.
2. Before first use and after each use, sterilize the above components.
Reprocessing Instructions
Preparation at the Point of Use:
1. Disconnect the silicone protective cover and handheld bracket before
cleaning. Remove gross contaminations from the components with code water
(<40°C) immediately after use. Don’t use a fixating detergent or hot water
(>40°C) as this can cause the fixation of residuals which may influence the
result of the reprocessing process.
2. Store the instruments in a humid surrounding.
WARNING
Do not submerge the components or wipe them with any of the following
functional water (acidic electrolyzed water, strong alkaline solution, or ozone
water), medical agents (glutaral, etc.), or any other special types of water or
commercial cleaning liquids. Such liquids may result in metal corrosion and
adhesion of the residual medical agents to the components.
Transportation:
Safe storage and transportation to the reprocessing area to avoid any damage
and contamination to the environment.
Preparation for Decontamination:
The devices must be reprocessed in a disassembled state.
WARNING
Observe suitable personal protective measures.
Pre-Cleaning:
Do a manual pre-cleaning, until the components are visually clean. Submerge
the components in a cleaning solution. Clean the surfaces with a soft bristol
brush.
Cleaning: