11 EMC Tables
Page 63 / 66
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare locations
where medical equipment and/or systems are used. When they are used in close proximity to
medical equipment and/or systems, the medical equipment and/or systems’ basic safety and
essential performance may be affected. The NanoPix1/NanoPix2 has been tested with the
immunity test level in the below table and meet the related requirements of IEC
60601-1-2:2014. The customer and/or user should help keep a minimum distance between
RF wireless communications equipment and the NanoPix1/NanoPix2 as recommended
below.
Test
freq
uenc
y
(MH
z
)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity
test level
(V/m)
385
380-390
TETRA 400
Pulse modulation
18Hz
1.8
0.3
27
450
430-470
GMRS 460
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band
13, 17
Pulse modulation
217Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
18Hz
2
0.3
28
870
930
1720
1700-19
90
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band
1,3,4,25;
UMTS
Pulse modulation
217Hz
2
0.3
28
1845
1970
2450
2400-25
70
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217Hz
2
0.3
28