Safety notes
Control unit
The control unit is classed as »conventional equipment« (closed equipment without protection against the ingress of water).
When using transmission settings 975, WS-75 and WI-75, Elcomed must be used exclusively with the following W&H-approved
surgical contra-angle handpieces: 975 AE, WS-75 E/KM, WS-75 LED G, WI-75 E/KM and WI-75 LED G. Use of other contra-angle
handpieces may result in deviation from the indicated torques and is the users responsibility. The specified transmission ratios
for the programs 1 to 6 must always be considered.
Mains cable
Only use the mains cable supplied.
Only connect to a grounded socket outlet.
Set up the device so that the power switch is easily accessible.
In dangerous situations, the device can be disconnected from the power supply using the power switch or mains cable.
The power switch can also be used to safely stop the device.
Power failure
In the event of a power failure, if the Elcomed is switched off, or when switching between programs, the last values set are saved and
re-activated when switching the device on again.
System failure
A total system failure does not constitute a critical fault.