Safety notes
Intermittent operating mode S3 (2min/10min)
The Elcomed is designed for intermittent operating mode S3 with an operating time of 2 minutes and an idle time of 10 minutes.
If the operating mode specified is observed no overheating of the system and therefore no injury to the patients, users or third
parties arises. The responsibility for the use and timely shutdown of the system lies with the user.
Coolant
The Elcomed is designed for use with physiological saline solution. Use only suitable irrigation fluids and follow the
manufacturer's medical data and instructions. Use the W&H irrigation tubing set or accessories approved by W&H.
You can purchase the coolant bottle or the coolant bag at a drugstore.
Sterility of irrigation tubing set
Sterile irrigation tubings are supplied with the equipment. These irrigation tubings are disposable articles and must be
discarded after each treatment! Please note the expiry date and the relevant regulations for disposal of irrigation tubings.
Only use disposable irrigation tubings with undamaged packaging.
Rotational energy
As a result of the rotational energy stored in the drive system – compared to the value set – fast deceleration of the bur can
cause the selected torque to be significantly overloaded, at times.
Observe the individual manufacturer's Instructions for use when adjusting superstructure screws. We would point out that
adjusting these screws with an electric motor presents a potential risk as described above.