EasyManua.ls Logo

Elcometer 311 - Taking a Reading; Measurement Procedure; Reading Interpretation and Data Transfer

Elcometer 311
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.6 SELECTING THE MEASUREMENT UNITS
The Elcometer 311 can take measurements in µm and mils. To set the
measurement units:
1 Press the ‘ softkey.
é
2 Use the softkeys to highlight ‘µm/mil’ and press ‘ ’ to
ê
alternate between ‘µm’ and ‘mil’, followed by the ç softkey to
return to the main reading screen.
www.elcometer.com
en-5
R
4 GETTING STARTED (continued)
5 TAKING A READING
Before taking a reading, it is good practice to check the performance
of the gauge using the calibration check procedure, see Section 7
‘Checking the Gauge Calibration’ on page en-6.
To take a reading:
1 Whilst holding the gauge perpendicular, gently place the probe
onto the surface.
4 The reading may be inaccurate if the probe is not held flat against the
surface.
2
The coating thickness is shown on the display.
4 ‘>500µm’ (‘>20mils’) indicates a reading outside the gauge range.
3 For subsequent readings, lift the probe off and then replace it on
to the coated surface.
®
Using ElcoMaster the Elcometer 311 Model T can simultaneously
transmit the live reading to a PC or mobile device for report generation,
see Section 8 ‘Downloading Data’ on page en-7 for more information.
The Elcometer 311 Model T calculates and displays statistical values
as each reading is taken, see Section 6 ‘Statistics’ on page en-6 for
details.

Related product manuals