23
elements
eONE - User’s guide
Figure 14
eONE Controls Voltage Protocols
Constant Voltage protocol
Figure 15
eONE Controls Voltage Protocols
Triangular wave protocol
Figure 16
eONE Controls Voltage Protocols
Seal test protocol
Figure 17
eONE Controls Voltage Protocols
Conductance test protocol
Constant voltage protocol:
The constant voltage is considered as a protocol with only the Vhold parame-
ter.
Triangular wave protocol:
It can be used to control the bilayer formation and it is used for the Capaci-
tance estimation feature (see details section 3.1.7).
Seal test protocol:
It is commonly used for setting up the sealing of the pipette to the cell in
patch clamp studies. Usually the Ne parameter is set to 0 to repeatedly apply
the stimulus to the cell.
Conductance test protocol:
It can be used to semi-automatically perform a channel’s conductance char-
acterization.