Safety
18
CAUTION
C12.0090
Risk of eye injury
Chemicals
When changing the chemicals, minute particles of chemicals may be
suspended in the air and cause burns to eyes.
Always wear protective goggles when working with chemicals.
Comply with the safety data sheets for the chemicals used.
CAUTION
C13.0000
Risk of injury
Pressurised system
The machine is pressurised during operation. The removal of the reagent
tube during operation results in an explosive escape of chemicals or
sample material and can cause injuries.
Never remove reagent tubes during operation.
CAUTION
C14.0094
Risk of burns
Hot catalyst furnace
The furnace temperature of the catalyst furnace can be up to 800 °C
during operation and cause burns.
Wear heat-resistant protective gloves when working with the catalyst
furnace .
Only replace the catalyst furnace filling after a sufficiently long
cooling period.