25
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV
contact discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV,
+/- 15 kV air
discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV
contact discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV,
+/- 15 kV
air discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-
4-8
30 A/m, 50
or 60 Hz
30 A/m, 50
or 60 Hz
It may be necessary to
position the breast pump
further from sources of
power frequency
magnetic fields or to install
magnetic shielding. The
power frequency magnetic
field should be measured
in the intended installation
location to assure that it is
suciently low.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment guidance
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test IEC 60601 test level Compliance level
Radiated
RF IEC
61000-4-3
10 V/m (minimum)
80 MHz –2,7 GHz
80 % AM at 1 kHz
10V/m (minimum)
Electromagnetic environment guidance: Portable and mobile RF communications
equipment should be used no closer to any part of the breast pump, including cables,
than the recommended separation distance calculated from the equation applicable to
the frequency of the transmitter. Recommend separation distance:
d = 1,2√P
d = 1,2√P 80 MHz - 800 MHz
d = 2,3√P 800 MHz - 2.7 GHz